
|
| |
Details
BLEEDING & TROUBLE SHOOTING OF CLOSE-LOOP SYSTEM
|
|
| |
A. Bleeding:
Bleeding
means removing air, trapped in the hydraulic circuit, i.e.
oil side of the intensifier, hydraulic hoses or piping and
cylinders.
Bleeding is necessary, first at the time of installation
and later when air gets trapped in the system.
If air is not bled properly, full pressure may not be achieved.
Also a jerky motion of the cylinders may be observed.
It is recommended to bleed the system from the highest point.
It is always better to have this highest point as an intensifier.
C.
Trouble shooting:
The intensifier will continue to work without any problem
unless there is any leakage in the system or there is air
in the system. Leakage in the system can be detected from
the oil level in the make-up oil bottle. This level should
remain constant.
1. Symptom -
No oil pressure from intensifier.
Cause - There is air in the system.
Remedy - Bleed the system to remove air.
See bleeding procedure.
2. Symptom - Oil level always decreases.
Cause - Oil is sucked into the system through
the make-up oil system, indicating leakages in the high
pressure line. Leakages may be in the connections or through
seals of cylinders.
Remedy - Stop leakages, change cylinder
seals, if necessary.
3. Symptom - Oil level increases in pressurised
mode.
Cause - In pressurised mode, make-up oil
system is cut-off from the high pressure side. A rise in
the oil level in the pressurised mode indicates that the
high pressure seal of the intensifier is leaking.
Remedy - Change the high pressure seal
of the intensifier.
4. Symptom - Oil level increases in unpressurised
mode.
Cause - There is air in the system. Due
to the compressibility of air, the trapped air gets compressed
in pressurised mode. In unpressurised mode, it expands
and pushes the oil to the make-up oil system. Hence the
oil level rises in the unpressurised mode.
Remedy - Bleed the system to remove air.
See bleeding procedure. Replace the oil if necessary.
^TOP
|
|
|